Message
Passwords do not match
Program memory error
Data memory error
No matching data found in memory
Sensor ID already exists, previous SN will
be overwritten
Temp. out of nLF correction range
8.5 Error limits
Message
Out of range, determine again
Buffer/standard temp. out of range
Offset out of range
Slope out of range
Wrong buffer
Invalid pH for user-defined buffer
ATC measured temperature is different to
the user-defined value
Conductivity measured temperature out of
range
SevenGo Duo pro™
Description and Resolution
The confirmation PIN does not match with the entered PIN.
Reenter PIN.
Meter recognizes internal error during start-up.
Switch the meter off and back on.
Contact METTLER TOLEDO service if the problem persists.
The data could not be stored into memory.
Switch the meter off and back on.
Contact METTLER TOLEDO service if the problem persists.
The entered filter criterion does not exist.
Enter a new filter criterion.
Two sensors with the same ID but different SN are not allowed in
the meter. If a different SN has been entered for this sensor ID
previously, the old SN will be overwritten.
Enter a different Sensor ID in order to keep the previous ID
and SN.
Conductivity measurements of natural water can only be
performed at temperatures from 0 ... 36°C.
Keep the sample temperature within the range.
Range not accepted
pH
mV
Conductivity
TDS
Salinity
Resistivity
T (pH)
T(cond.)
| Eref1-Eb | > 60 mV
| Eref1-Eb | > 60 mV
I ∆Eref1I < 10 mV
I ∆pHI < 1 pH
| tATC-tbuffer | > 1 °C
T: < -5 °C or > 105 °C
<-2.000 or > 19.999
<-1999.9 or > 1999.9
< 0.00 µS/cm or > 1000 mS/
cm
< 0.00 mg/L or > 600 g/L
< 0.00 ppt or > 80.0 ppt
< 0.00 MΩ•cm or > 100.0
MΩ•cm
< 5 or > 50 °C
< 0 °C or > 35 °C