Technical Specifications
Description
Physical Properties
Belt length
Belt material
Cable length
Cable material
Module
Battery
Battery Life
Enviromental Specifications
Temperature
Relative Humidity
Pressure
Output Specifications
Output Signal
Frequency Range
Sensitivity
Polygraph Settings
Sensitivity
Low Frequency Filter /
Time Constant
High Frequency Filter
Properties
Adjustable 24"-79" (60-200cm)
Elastic band with Insulated wire
80" (205cm)
PVC jacketed; ZIP-style cord
Box: ABSCable: PVC jacketed; Zip Style cord
3V Lithium (integrated in the device)
The battery life is estimated to be 3000 hours or 375 eight hour studies. The
battery is activated once the sensor module is connected to the belt.
Operation : 40°F to 120°F (+5°C to +50°C)
Storage: 0°F to 120°F (-20°C to +50°C)
Operation : 15-95% (non-condensing)
Storage: 10-95% (non-condensing)
Withstands atmospheric pressures from 7.3 psi to 29 psi
Maximum signal amplitude: ±5mV
0.2 to 3 Hz
Approximately 50μV/mm
Adjustment of the sensitivity up or down is typically required. Response is
dependent upon variables such as sensor application and patient effort.
0.16 Hz (or 1 second or longer) Shorter time constants or higher low
frequency filter settings will significantly attenuate waveforms
35 Hz.
7