OPERATION IS ONLY PERMITTED WHEN OPERATIONAL AND PROCESS-RELATED ELECTROSTATIC CHARGES ARE ELIMINATED. USE THE
EQUIPMENT ONLY WHEN ELECTROSTATIC PROCESSES LEADING TO PROPAGATION BRUSH DISCHARGE IS IMPOSSIBLE. KEEP THE
EQUIPMENT AWAY FROM PROCESSES THAT GENERATE HIGH CHARGING POTENTIAL SUCH AS ELECTROSTATIC COATING, RAPID TRANSFER
OF NON-CONDUCTIVE MATERIALS, RAPID AIR JETS, AND HIGH-PRESSURE AEROSOLS.
THE NON-METALLIC PARTS (DISPLAY AREA OF IND500x) INCORPORATED IN THE ENCLOSURE OF THIS EQUIPMENT MAY GENERATE AN
IGNITION-CAPABLE LEVEL OF ELECTROSTATIC CHARGE.
a) THE EQUIPMENT SHALL NOT BE INSTALLED IN A LOCATION WHERE BUILD-UP OF ELECTROSTATIC CHARGE ON SUCH SURFACES SHALL OCCUR.
b) THE DISPLAY AREA OF THE EQUIPMENT SHALL ONLY BE GENTLY CLEANED WITH A DAMP CLOTH.
c) STRONG SOLVENTS ARE NOT ALLOWED TO BE USED FOR CLEANING OF THE WHOLE ESD PROTECTED PLASTIC KEYPAD.
d) THE ENTIRE PLASTIC KEYPAD MUST BE REPLACED WHEN THE ESD PROTECTION LAYER IS VISIBLY WORN.
THE MODEL IND500x WEIGHING INDICATOR CIRCUITS SHALL BE LIMITED TO OVERVOLTAGE CATEGORY I OR II AS DEFINED IN IEC 60664-1.
THE MODEL IND500x WEIGHING INDICATOR SHALL NOT BE USED WHERE UV LIGHT OR RADIATION MAY IMPINGE ON THE ENCLOSURE.
Specifications and Operator Interface
Specifications
Enclosure Types
Dimensions (l × w × d)
Shipping Weight
Environmental Protection
Storage Environment
Operating Environment
Hazardous Areas
Power
Display
Weight Display
Scale Types
Number of Analog Load Cells Four 350 Ohms, 1-3 mV/V
Analog/Digital Update Rates
Interface and Function Update
Rates (max)
Clock Accuracy
Load Cell Excitation Voltage
Minimum Sensitivity
Keypad
30753869
|
B
|
01/2023
Harsh environment desk/wall/column-mount: Type 304 stainless steel enclosure
Harsh environment: 289 mm × 184 mm × 162 mm (7.2 in x 11.4 in. x 6.4 in.)
4.0 kg (8.8 lb)
Harsh environment enclosure meets IP65 requirements.
-20°C to 60° C (-4° to 140°F), 10% to 95% relative humidity, noncondensing.
-10° to 40° C (14° to 104°F), 10% to 95% relative humidity, noncondensing.
The IND500x is approved for use in Division 1 and Zone 1/21 hazardous areas.
APS768x power supply with intrinsically safe outputs.
4.3" TFT color display (480 x 272)
Displayed resolution of 100,000 counts for analog load cell scales.
Display resolution for high-precision IDNet and SICSpro bases is determined by the specific base used.
Analog load cells, High-Precision IDNet, High-Precision SICSPro
Internal: Analog: >366 Hz; IDNet: determined by base; SICSpro: 50 Hz for bases that include Advanced Setup
Mode. Target Comparison: up to 50 Hz
Weight display: 10 Hz
Internal discrete I/O: 50 Hz
External discrete I/O (ARM100): 25 Hz
PLC cyclic data: 25 Hz
SICS continuous (SIR): up to 20 Hz
MT Continuous Output: up to 20 Hz
Continuous Template (serial): 18 Hz (115.2Kbaud), 14 Hz (38.4Kbaud)
Continuous Template (Eprint): 20 Hz
< 1 second / day (without time server access) at 25° C consistent room temperature.
4.8 VDC
0.3μV/e approved
27 keys; polyester overlay (PET) with hardcoating, Polycarbonate (PC) display lens
WARNINGS
METTLER TOLEDO IND500x Quick Guide
3