EMC Guidance
The ME EQUIPMENT or ME SYSTEM is suitable for home
healthcare environments
Warning: Don't use near active HF surgical equipment
and the RF shielded room of an ME system for magnetic
resonance imaging, where the intensity of EM disturbances
is high.
Warning: Use of this equipment adjacent to or stacked with
other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that
they are operating normally.
Warning: Use of accessories, transducers and cables other
than those specified or provided by the manufacturer of
this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this
equipment and result in improper operation."
Warning: Portable RF communications equipment (including
peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any
part of the equipment , including cables specified by the
manufacturer. Otherwise, degradation of the performance of
this equipment could result.
Technical description:
1. All necessary instructions for maintaining BASIC
SAFETY and ESSENTIAL PERFORMANCE with regard
to electromagnetic disturbances for the excepted
service life.
2. Guidance and manufacturer's declaration
-electromagnetic emissions and Immunity
Table 1
Guidance and manufacturer's declaration -
electromagnetic emissions
Emissions test
RF emissions, CISPR 11
RF emissions, CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions, IEC 61000-3-3
24
Compliance
Group 1,
Class B
Not applicable
Not applicable