Dräger X-act 7000 Manual Tecnico página 55

Ocultar thumbs Ver también para X-act 7000:

Publicidad

Idiomas disponibles

Idiomas disponibles

Error
Error
524332 -
self-test
failed
Error
524333 -
self-test
failed
Error
524336 -
self-test
failed
Error
2097154 -
self-test
failed
5.3
Warnings
Warnings
X-act 7000 - low battery
Measurement - aborted
Service - expired
MicroTubes - used up
MicroTubes - not found
|
Technical Manual
Cause
The MicroTubes intake/discharge
could not be correctly moved during
the self-test
The pneumatics system could not
generate the necessary mass flow
during the self-test
The camera could not be correctly
accessed during the self-test
The component driver could not be
correctly initialized during the self-
test
®
X-act
7000
Cause
The battery charge of the
X-act 7000 is low
User has manually
aborted the measurement
Service interval exceeded Servicing of the X-act
All 10 measurements of
the inserted MicroTubes
have been used up
The RFID tag of the Micro-
Tubes was recognized, but
the MicroTubes could not
be drawn in
Troubleshooting
Remedy
Make sure that the wells/doors
for MicroTubes are not blocked;
re-start the device and carry out
the process once again
make sure that the gas outlet is
not sealed (e.g. by fingers, con-
tamination)
Repeat the self-test
Repeat the self-test
Remedy
Promptly insert new bat-
teries in the X-act 7000
(see IFU)
Start a new measurement
if necessary
7000 by DrägerService is
recommended
Use new MicroTubes for
further measurements
Check whether the Micro-
Tubes were inserted far
enough into the X-act
7000 for them to be cap-
tured by the automatic
intake; if necessary, push
the MicroTubes in a bit
faster and with slightly
more pressure; if no mea-
surement is intended,
remove the MicroTubes
from the immediate vicinity
of the X-act 7000
|
en
55