Descargar Imprimir esta página

Steinberg Systems SBS-BS-5000 Manual De Instrucciones página 27

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 85
3.3.8
Value memory function (HOLD)
This function is used to temporarily store the measured value. To use
this function:
3.3.9
Calibration
The equipment is shipped as calibrated at the factory and calibration
is usually not required unless there are major inaccuracies in the
weighing results or the equipment has been moved to work away
from its original location or has been serviced. For calibration, you
need a weight or weights (not included in the device kit) of at least
2/3 of the maximum allowable load of the device. Before performing
the initial calibration, warm up the device for approx. 30 minutes (see
more in "ASSEMBLY OF THE DEVICE").
To calibrate the device:
Press the HOLD button again, and the display will show "Ani.
Fd." Then press the UNIT button to enter the selection of the
sensitivity/accuracy of the measurement. Use the ZERO
button select the range from 0.5, 3, 5 or 10 - confirm the
selection with the HOLD button. It is a setting of the scale
unit in the range of 0.5, 3, 5 or 10 that the device shows on
the display. The smaller the unit, the more accurate the
measurement, but the less stable the result. The larger the
scale is, the less accurate result but the more stable
measurement.
Pressing the HOLD button again will trigger the message
"Ani. Ft." Then press the UNIT button to enter the setting of
the time for displaying the weighing result and use the ZERO
button to select from the range 1, 2, 3, or 4 (seconds). Use
the HOLD button to confirm the selected setting.
Turn on the device with the ON/OFF button and place the
item on it.
Wait for the weighed value to stabilize and then press the
HOLD button until "." appear.s
Remove the weighed item from the device and the weighed
value will continue to be displayed.
To delete the stored value, press the HOLD button again.
Press and hold the HOLD button and press the ON/OFF
button to turn on the unit. The display will show "CAPu =".
Press the HOLD button again, the display will show "CAP."
Then press HOLD again and the message "CALu =" will
appear. Press the UNIT button to select the unit of
measurement to be calibrated (kg or lb), which will be

Publicidad

loading

Este manual también es adecuado para:

Sbs-bs-2500