Determining/Setting the cell constants
C
A L
1/
cm
0 7
4
3
Tref25
Cal
C L
E
L
6
8 0
S/
1/
0 1
6
4
Tref25
ARng
nLF
C L
E
L
9
2 5
S/
1/
0 1
0
0
Tref25
ARng
nLF
8
Press <RUN/ENTER>. The AutoRead measurement begins.
If the measured value is stable, the instrument displays the determined
cell constants and the calibration evaluation. The measuring instrument
automatically stores the cell constants.
Switch to the measuring mode with <M>.
Procedure for manually setting the cell constant:
Press the <CAL> key repeatedly until the
Then press the <RUN/ENTER> key.
Press the <CAL> key until a measured value and a value for the cell con-
stant appears in the required setting range.
cm
Set the value of the cell constant with <L> <M>.
cm
Switch to the measuring mode with <M>.
Procedure for selecting fixed cell constants 0.010 cm
Press the <CAL> key repeatedly until the
Then press the <RUN/ENTER> key.
Press the <CAL> key until a measured value and the cell constant
-1
0.010 cm
cm
Switch to the measuring mode with <M>.
cm
appear.
Cond 3300i/3400i
CELL display appears.
-1
:
CELL display appears.