Procedure:
1. It is recommended to offset the sensor outside (you
can get it out through a window if it is connected to a
USB module).
2. The CO
2
offset. Connect the CO
 USB-200 module for On-line experiments.
 BAT-200 battery module for Off-line experiments.
 VIEW-101 graphic display module and BAT-200
battery module for use with the NeuLog Viewer.
3. After the sensor has been connected to a power
supply for at least 30 minutes, check that the values
are relatively stable.
4. The sensor now needs to be offset – hold the probe
vertically with the opening facing the ground.
5. Press and hold the blue "Start/Stop" button on the
faceplate of the sensor for 3 seconds.
6. The CO
Sensor is now offset for use.
2
V2018.01
sensor needs a power supply to properly
20
sensor to:
2