SPARE PARTS
Please contact your local sales specialist or email:
cpspares@antylia.com to enquire about available spare parts
Routine Maintenance
To maintain the efficiency of your BS-200 we recomend that every
six months you carry out the following maintenance:
WARNING: Ensure the equipment is disconnected from the power
supply before attempting any maintenance, servicing or cleaning.
Remove the screws from the sides of the unit; unscrew the
location bosses and remove the cover.
Remove the drive ring and check it for wear or deterioration.
Unscrew the shafts from the pulleys. Clean the shafts and the
bearings in the pulleys with alchohol or meths. DO NOT ADD ANY
LUBRICANT.
Replace the pulleys and shafts then replace the drive ring.
Replace the cover and screws, then fit the appropriate location
bosses.