ORAL USE (FIGURE 1)
· The mouth should remain closed 3 to 5 minutes before attempting a reading.
· Place the thermometer in the mouth under the tongue so that it rests to the left or right
of the root of the tongue.
· Use downward tongue pressure to hold the thermometer in place.
· Hold the thermometer to keep it from sliding around in the mouth.
· When the reading is ready, the thermometer will emit a beep sound, remove
the thermometer and read the result.
UNDERRARM USE (FIGURE 2)
· The underarm should remain clamped 3 to 5 minutes before attempting a reading.
· Place the probe tip in the center of the underarm.
·
Lock the probe tip under the arm, using the arm to slightly apply pressure inwards.
·
When the reading is ready, the thermometer will emit a beep sound, remove
the thermometer and read the result.
CARE AND MAINTENANCE
· Wipe the thermometer with a soft clean cloth before and after each use.
· Never put the thermometer into any liquid.
· Do not shock, knock or move the thermometer when using it.
· The thermometer is not a toy; please keep it out of the reach of children.
· Adult supervision is required when taking a temperature on children.
· Contact your physician for specific information about your temperature readings.
·
Store the thermometer in a clean, safe and dry location. Do not store the thermometer where it will be exposed
to direct sunlight, dust, etc. Keep it far from heat source. Storage temperature: -10°C~+60°C; Storage humidity <95%
·
Do not use a mobile phone near the thermometer or in places where strong static electricity or electromagnetic field
at present.
SPECIFICATIONS
Measurement Range
Measurement Accuracy
Temperature Display
Resolution
Buzzer Notification
Memory Function
Usage Environment
Storage Environment
Battery
Reference Standard
32.0°C~43.9°C (90.0°F~111.0°F)
"Lo"displayed when the thermometer of the sensing section is less than 32.0°C (90.0°F)
"Hi" displayed when the thermometer of the sensing section is greater than 43.9°C (111.0°F)
±0.1°C (35.0°C~41.0°C)
±0.2°F (95.0°F~105.8°F)
3.5-digit LCD display
0.1°C (0.1°F)
Press ON/OFF button or the temperature measurement is complete; the buzzer will emit
beep sound.
The last reading will automatically be stored in memory.
Surrounding Temperature: 10°C~40°C (50°F~104°F), Relative Humidity: 30%RH~85%RH
In the 16°C~31°C environment temperature range, measurement can be done in 8 to 10
seconds. Otherwise it will need 20 seconds to 2 minutes to complete temperature measu-
rement.
Surrounding Temperature: -10°C~+60°C (14°F~140°F), Relative Humidity: 10%RH~95%RH
LR41
EN12470-3: 2000; clinical thermometers ASTM E-1112; electronic thermometers
ENGLISH