Descargar Imprimir esta página

Aqua HC 100 pH Manual De Instrucciones página 9

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 14
EXAMPLES OF PROGRAMMATION
MANUAL 100 pulse/min
Press the F button and select the F1 function. Press Enter and while the display is blinking select 100 by
pressing Up e Down, press Enter to confirm, the green Led will light up and the pump will work.
PROPORTIONAL to the pH measure with a set-point of 7,4pH, maximum flow for an pH value of 9pH
and maximum number of injection per minute 80.
After having calibrated the probe, choose the pH set-point, the extreme value (L1) of the proportional
interval and the maximum number of injections per minute as follows:
Press the F button until the function Set – set-point function.
Press the Enter button to confirm the choice. Modify the reading pressing the UP and DOWN buttons until
the value of 7,4 pH and then press Enter to confirm.
Press the F button until the function L1 – proportional function- will appear.
Press Enter to confirm the choice. Set the value 9 pH by pressing the UP and DOWN buttons and confirm by
Enter. Press the F button until the function F2 – maximum proportional flow function- will appear.
Press Enter to confirm the choice. Set the value 80 by pressing the UP and DOWN buttons and confirm by
Enter.
The Led will light up and the pump will work.
17/11/2014
ADSP7000180A Rev.2.2
9

Publicidad

loading

Este manual también es adecuado para:

Hc 797 ph