Battery Replacement (See Fig. 4); Fuse Replacement; Specifications - Amprobe RS-3 PRO Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 44
or by other qualified instrument service personnel. The front panel and case
can be cleaned with a mild solution of detergent and water. Apply sparingly
with a soft cloth and allow to dry completely before using. Do not use
aromatic hydrocarbons or chlorinated solvents for cleaning.

battery Replacement (see fig. 4)

 Warning
To prevent electrical shock or meter damage, disconnect the meter's test
leads from any circuit and the meter, then turn the meter off before
removing the battery cover. battery replacement should be performed in a
clean environment and with appropriate care taken to avoid contaminating
the meter's interior components.
1.
Remove the screw and lift the battery cover.
2.
Remove the battery using the pull strap.
3.
Replace the battery with the same type, 1.5 V AAA battery (NEDA-24A, IEC
LR03). Note polarity of the battery.
4.
Replace the battery cover and screw.

fuse Replacement

1.
Remove the screw and lift the battery cover.
2.
Remove the battery using the pull strap.
3.
Remove the fuse using the pull strap.
4.
Replace the fuse with F 0.5 A Fuse (FP500)
5.
Replace the battery noting the polarity of the battery.
6.
Replace the battery cover and screw.

SPECIfICATIONS

general
Display: Analog scales with clamping needle
Power Supply:
Volts / amps: powered by circuit under test
Ohms: 1.5 V AAA battery (ANSI/NEDA-24A, IEC LR03)
Low Battery: Below approx. 1.2 V
Environmental: Indoor operation; below 2000m (6,562ft)
Operating Temperature : -15°C to 50°C (5°F to 122°F) ; < 80%
Storage Temperature: -20ºC to 60ºC (-4ºF to 140ºF), < 80% R.H. (with battery
removed)
Temperature Coefficient: nominal 0.15 x (specified accuracy)/ºC @ 0ºC to 18ºC
or 28ºC to 40ºC (32°F to 64°F or 82°F to 104°F)
Dimension:
RS-1007 Pro: 261 x 102 x 41 mm (10.3 x 4.0 x 1.6 in.)
RS-3 Pro: 243 x 102 x 41 mm (10.3 x 4.0 x 1.6 in.)
Weight:
RS-1007 Pro: 0.474 kg (1.05 lb.)
RS-3 Pro: 0.455 kg (1.0 lb.)
Jaw opening / Conductor diameter:
RS-1007 PRO: 47mm (1.88") max
RS-3 PRO: 41mm (1.64") max
Safety LVD: Meets EN60101-1:2001; EN61010-2-032(2002), Category IV - 600
Volts ac; pollution degree : II; class 2
 EMC: EN 61326-1:2006 This product complies with requirements of the
following European Community Directives: 2004/108/EC (Electromagnetic
Compatibility) and 2006/95/EC (Low Voltage) as amended by 93/68/EEC (CE
Marking). However, electrical noise or intense electromagnetic fields in the
vicinity of the equipment may disturb the measurement circuit. Measuring
instruments will also respond to unwanted signals that may be present within
the measurement circuit. Users should exercise care and take appropriate
precautions to avoid misleading results when making measurements in the
presence of electronic interference.
Electrical
23ºC ± 5ºC (73ºf ± 9ºf) <75%RH
AC Voltage (50/60 Hz)
Ranges: 150 / 300 / 600 Vrms
Accuracy: ± 3% of Full Scale.
Input Impedance: 1.5 MΩ
7

Publicidad

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

Rs-1007 pro

Tabla de contenido