5.7. Memory – displaying the last 99 measurements
The measured results are stored in the instrument until a new
measurement is carried out. With the unit in the OFF position,
press the MEMORY button twice. The screen will show the last reading along with time
and date. The MR symbol is displayed in the lower left. Each time you press the memory
button an earlier measurement will be displayed along with time and date.
If an irregular heartbeat was detected in any measurements
stored in memory, the
To clear all of the stored memories, press and hold the memory
button until the CL symbol appears on the screen.
5.8. Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for any reason
(e.g the patient feels unwell), the ON/OFF button can be pressed at any
time. The device then immediately lowers the cuff pressure automatically.
6. Error Messages/Troubleshooting
If an error occurs during a measurement, the measurement is discontin-
ued and a corresponding error code is displayed. We recommend you
switch the device off and take a new measurement making sure you
allow time for relaxation.
(example: error no. 1)
Error No.
ERR 1
ERR 2
4/5/06
9:36 AM
icon will be displayed.
Possible cause(s) / Solutions
The tube may have loosened, or no pulse
was detected.
*Ensure cuff connections are tight with
proper cuff placement. See section (5.3)
Unnatural pressure impulses. Reason:
The arm was moved during the
measurement (Artefact).
Repeat measurement keeping arm still.
Page 13
Last
reading
14 readings
in memory
14
MR
MR
12