2.8. Sleeping
2.8.1. How into the Sleeping State
1. Put the handle on the handle holder, the system will to the sleeping
state come automatically.
2. During the system from work state to sleeping state, the window will
display "Off". and then when the temperature cooling down to 100℃,
the unit comes into the sleeping state. In the sleeping state, the LCD
will display "Sleep".
2.8.2. Resume
1. When the handle is putting on the handle, it cannot be resumed from
sleeping.
2. Take down the handle from the holder. At the time, the unit will work
at once.
2.9. Calibration
It is better to calibrate the temperature after changing the heater element.
Adopt two point temperature calibrate, the method of calibrating the
temperature is as following:
1. In work state, set the calibration temperature 300℃ and 500℃ respective
calibrate.
When the temperature is stable and the window displays "Real Temp", test
the outlet temperature of the handle with thermometer and write down it.
2. Into the calibration mode: When the temperature is stable, press the keys
"CH1" and "CH3" keys at the same time about 15seconds, and then the LCD
will display "cal" about two seconds and the system will enter into the
temperature calibrating mode.
Input the testing temperature after the hundreds digit twinkling.
3. Press "TEMP▲"or "TEMP▼" key to setting the data and press "AIR
▲"or "AIR▼" key to move the cursor. After inputting the testing result,
click the "CH2" key.
9
USER MANUAL