Battery Maintenance
You should routinely check the battery's charge status . Carefully monitor batteries
that are approaching the end of their estimated life . Consider shipping the device
back to DJO to replace the battery if you note either of the following conditions:
1 . The battery run time drops below about 80% of the original run time .
2 . The battery charge time increases significantly .
If a battery is stored or otherwise unused for an extended period, be sure to follow
the storage instructions in this document .
Storage
Charge or discharge the battery to approximately 50% of capacity before storage .
Charge the battery to approximately 50% of capacity at least once every three
months .
The battery is best stored at 23±5°C . For periods of less than 1 month, the battery
can be stored between -25°C to 70°C .
NOTE: The battery self-discharges during storage . Higher temperatures (above 20°C
or 68°F) reduce the battery storage life .
Handling Precautions
Do not disassemble, crush, or puncture a battery .
Do not short the external contacts on a battery .
Do not dispose of a battery in fire or water .
Do not expose a battery to temperatures above 70°C (158°F) .
Keep the battery away from children .
Avoid exposing the battery to excessive shock or vibration .
Do not use a damaged battery .
If a battery pack has leaking fluids, do not touch any fluids .
Properly dispose of a leaking battery pack .
In case of eye contact with fluid, do not rub eyes . Immediately flush eyes
thoroughly with water for at least 15 minutes, lifting upper and lower lids, until no
evidence of the fluid remains . Seek medical attention .