EN-16
5.3
Electrode maintenance
Make sure the electrode is filled with electrolyte solution. Always store the electrode according to
the electrode instruction manuals and do not allow it to dry out.
If the electrode response becomes sluggish or the slope is not acceptable, try the following:
o
Soak the electrode in 0.1M HCl for at least 8 hours.
o
For fat or oil contaminant, degrease the membrane with cotton wool soaked in either
acetone or a soap solution.
After electrode treatment, a new calibration should be performed. If the electrode slope is still not
acceptable, the electrode might need to be replaced.
5.4
Self-diagnosis
When the meter is on, press and hold
meter displays the full screen. Each icon blinks one after the other.
This way you may check whether all icons are correctly shown. The next step is to check that the
keys are functioning correctly. This requires user interaction.
When b blinks, five icons are displayed.
Press the five keys in any order. Each time you press a key an icon disappears from the
screen, continue to press the other keys until all the icons have disappeared.
When the self-diagnosis has been completed successfully, PAS appears. If self-diagnosis fails,
error message Err 1 appears.
Note: You have to finish pressing all five keys within 2 minutes, otherwise Err 1 appears and you
will have to repeat the procedure.
5.5
Recover factory settings
When the meter is off, press and hold
seconds, the screen displays
Press
then restart the meter.
Or press
and blinks, this means "Reset". There are 2 options:
to reset factory settings (MTC, slope and offset, etc.), display
to quit the setting, display
STARTER 3100 Bench pH Meter
and
&
then turn off the meter.
simultaneously until the
&
together for 3