Descargar Imprimir esta página

Inventis FLUTE & VIOLA Guia De Inicio Rapido página 27

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 63
start the test (START/STOP button). The test will start only if the status
of the probe is indicated as READY;
a first tympanogram is generated and the pressure in the ear canal is
raised automatically to +400 daPa;
invite the patient to swallow several times, as indicated in the pop-up
message, and press the CONTINUOUS* function button;
a second tympanogram is generated and the pressure in the ear canal is
lowered automatically to –400 daPa;
invite the patient to swallow several times, as indicated in the pop-up
message, and press the CONTINUOUS* function button;
a third tympanogram is generated and all three graphs are shown in the
display.
Eustachian Tube Function test (ETF) with perforated
eardrum
Select the ear being examined, left side or right side.
Select the maximum pressure (PRESS. MAX* function buttons), or,
the pressure that will be established in the middle ear of the patient for
the purposes of assessing Eustachian tube function;
select the duration of the test by pressing the DURATION* function
button (selectable values are 20, 30, 40 or 50 seconds)
start the test (START/STOP button). The test will start only if the status
of the probe is indicated as READY.
ask the patient to swallow so that a variation of pressure can be induced
in the examined ear, and functionality of the Eustachian tube thus
evaluated.
Quick A / Quick B
Quick A and Quick B are sequential tests in which the tympanometry
examination and then the acoustic reflex examination (configurable on the
Settings page of the instrument) are run automatically one after the other.
access the window relative to the tympanometry examination;
select the ear being examined, left side or right side;
select the rate at which the pressure sweep occurs during the course of
the test (P. RATE *);
start the test (START/STOP button). The test will start only if the status
of the probe is indicated as READY.
once the test has been concluded, press the scroll button (<<* or >>*)
to change the type of test displayed.
Pure-tone audiometry
select the type of threshold to examine (HL/UCL*);
select the stimulus input signal (21):
select the masking input signal (22):
(Viola only)
;

Publicidad

loading

Productos relacionados para Inventis FLUTE & VIOLA