Do not realise measurements if the battery compartment is
open.
In order to realise continuity test it is necessary to realise the
following actions:
place the rotational selector in the position Ω
connect the red test lead to sockets V/Ω/TEMP and the black
one to sockets COM,
press the MODE button until
touch the test probes to the diode under test - forward voltage
will indicate 0,4V to 0,7V, reverse voltage will indicate "OL";
shorted devices will indicate near 0mV and an open device will
indicate "OL" in both polarities,
having done the test disconnect the test leads from the meter.
5.6
Frequency or % duty cycle measure-
ments
Do not realise measurements if the battery compartment is
open.
In order to realise frequency or % duty cycle measurement it is
necessary to realise the following actions:
place the rotational selector in the position V ,
connect the red test lead to sockets V/Ω/TEMP and the black
one to sockets COM,
select Hz or % duty with the Hz/% button,
touch the test probe tips to the circuit under test,
read the frequency on the display,
having done the measurement disconnect the test leads from
the meter.
WARNING:
WARNING:
CMP-400 – USER MANUAL
appears on the display,
,
43