Appendix EMC Declaration
The equipment is intended for use in the electromagnetic
environment specified below.
The customer or the user of the EQUIPMENT should assure
that it is used in such an environment.
The Transcutaneous Electrical Nerve Stimulators is suitable
for use in a professional health care environment, not
including areas where there are sensitive equipment or
sources of intense electromagnetic disturbances, such as the
RF shielded room of an imaging system magnetic resonance
imaging, in operating rooms near active AF surgical
equipment, electrophysiology laboratories, armored rooms
or areas where short wave therapy equipment is used.
• Do not use the system around strong electric filed,
lectromagnetic filed (e.g. MRI scan room) and mobile
wireless communication devices. Using the device in
an improper environment may cause malfunction or
damage.
• The compliance with EMC and EMI regulation cannot
be guaranteed by the use of modified cables or those
which does not comply with the same standards under
what the equipment was validated.
• The system must not be used adjacent or supported by
other equipment. The recommendations of this manual
must be followed.
• Do not use accessories, transducers, internal parts
of components and other cables other than those
previously specified by the manufacturer. This
may result in increased emission or decreased
electromagnetic immunity and result in improper
operation.
• Portable RF communications equipment (including
peripherals such as antenna cables and external
antennas) should by used no closer than 30cm to any
16