H-2382, H-2383
H-2384
SCOUT PRO BALANCE SCALE
ERROR CODE LIST
Err 1 chEct
Invalid checksum data.
Err 2 LoAd
Overload (>cap+90d) or Underload.
Err 4 WEIght
Incorrect calibration weight.
Err 5 999999
Displayed value>99999.
Err 6 count
Parts counting error – balance exits parts
counting.
Err 8 FULL
RS232 buffer is full.
Err 9 dAtA
Internal data error. The next button
press or a 5 second time-out causes the
balance to turn off.
PAGE 2 OF 6
1-800-295-5510
uline.com
ERROR CODES
RECOMMENDATIONS
1. Cycle balance ON/OFF
2. Return balance for service
1. Remove load
2. Make sure shipping lock is unlocked
3. Recalibrate
1. Use correct calibration weights
2. Check environmental conditions
1. Exit application
2. Rezero balance
1. Exit application
2. Rezero balance
1. Cycle ON/OFF
2. Check RS-232 settings in balance and computer
1. Cycle ON/OFF
2. Return balance for servicing