Do not point the measuring end of the sensor up or damage to the sensor can occur (e.g., during storage,
transport, calibration, installation and operation).
Keep the sensor at 0 to 40 °C (32 to 104 °F) or damage to the sensor will occur.
The GS1440 and GS2440EX sensors continuously measure the hydrogen sulfide (H
concentration in liquids and air. Refer to
The sensor is used with a device that supplies 4–20 mA loop power to the sensor and receives the
sensor measurements as a 4–20 mA analog signal or RS-232 digital signal.
The GS1440 is made for use in non-hazardous locations. The GS2440EX is made for use in
hazardous locations.
Figure 1 GS1440 and GS2440EX
1 Membrane
2 Sensor cap
3
3 Status LED
(Table
4)
4 Attachment hook
Color
Green, one flash
Green, two flashes
Green and red flashes
Red, flashes
3
The status LED continuously flashes while power is supplied to the sensor.
8 English
N O T I C E
Figure
Table 4 Status LED
Status
The 4–20 mA analog output represents 0–5 mg/L (liquids).
The 4–20 mA analog output represents 0–1000 ppm (air).
The sensor is in calibration/configuration mode.
Temporary (with calibration cap attached): Calibration was not successful.
Continuous (without calibration cap attached): There is a problem with the sensor.
1.
5 RS-232 connector
6 Grounding screw (M6)
7 Power/4–20 mA connector
S)