Descargar Imprimir esta página

ivory Neo Line SL211E Instrucciones De Uso página 41

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

CAUTION!
Incorrect or inadequate spindle cleaning can cause the
spindle to loosen and accidentally release the bur during
functioning.
CAUTION!
After lubricating, hold the handpiece with the chuck
facing down. Operate the handpiece for 30 seconds to
expel any excess oil. If you notice any dirt coming out,
repeat the entire cleaning and lubrication process.
-Sterilisation
Sterilisation in an autoclave conforming to: EN 13060,
EN ISO 17665-1
CAUTION!
- Follow the instructions of the autoclave manufacturer.
- Clean, disinfect and lubricate the handpiece before
sterilisation. The presence of blood residues on the
external or internal surfaces can lead to the formation
of corrosion.
- Make sure that the handpiece is properly dried after
sterilisation.
Authorised sterilisation procedures
1) Seal the handpiece and its accessories in sterilisation
bags conforming to EN 868-5.
2) Sterilise the bagged handpiece in an autoclave
according to EN ISO 17665-1 standard and respect the
following conditions:
- Class B steam sterilisation in autoclaves conforming to
EN 13060 standard. Sterilisation time at least 3.5 minutes
at 135 °C.
3) Remove the bag from the autoclave.
39
Eng

Publicidad

loading