Frontal Panel; Operation Mode; Switching On; Neutral Electrode's Circuit - Gima DIATERMO 108 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

DIATERMO 108
GIMA SpA
MA254IGBEa

4.2 FRONTAL PANEL

1 Mains power-on light.
2 Foot-switch connector.
3 Output power display.
4 Key for output power decrease and/or increase.
5 Function keyboard.
6 OC (open circuit) alarm.
7 Return plate connection.
8 Handle connection.

4.3 OPERATION MODE

4.3.1 S
O
WITCHING
N
When switched on the electrosurgical unit performs automatically a test to establish the correct
operation of itself. In case anomaly is found it is shown coded according to the chart codes brought
in the chapter MAINTENANCE.
At the end of the control the equipment restores last use operational conditions.
4.3.2 N
E
'
C
EUTRAL
LECTRODE
S
IRCUIT
The neutral electrode is continually watched by a special circuit that prevents, danger of burns to the
patient due the loss of contact between the reference plate and the patient skin, if split electrode is
used. The circuit also watched the connection of the plate to the unit.
If the impedance value of the patient circuit is more than 200 ohm the value is not accepted.
In order to reduce the acoustic pollution, the sound alarm is present only when pressed the foot-
switch.
Manuale d'Istruzioni / Instruction's Manual / Manual de Instrucciones
38