GIMA SpA
•
Check of the protections against the danger due to liquid's pouring, dripping, moisture, liquid's
penetration, cleanliness, sterilization and disinfection.
•
Check of the Equipment's Datas on the Label.
•
Check of the availability of the Instruction's Manual.
•
Check the functioning of the H.F. output controls.
•
Check the uniformity of the resistance through the surface of the patient plate.
•
Test the earth conductivity resistance.
•
Test the earth leakage current.
•
Test H.F. leakage current.
•
Control of the neuromuscular stimulation.
•
Control of the accuracy of the output power.
Manuale d'Istruzioni / Instruction's Manual / Manual de Instrucciones
DIATERMO 108
MA254IGBEa
45