Repairs; Fuse Substitution; Checking Of The Equipment Before Each Use; Function And Safety Check And Test - Gima DIATERMO 108 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

GIMA SpA
Problems
Error Code 95
Error Code 99

6.5 REPAIRS

High frequency cables and electrode holder handle cannot be repaired. Always substitute a damaged
part with a new one.
6.5.1 F
S
USE
UBSTITUTION
Before substituting the fuse, disconnect the unit from the mains system.
Only use fuse of the kind 5x20; they must have those characteristics: T2A (slow) (230Vac mains
voltage), T4A (115Vac mains voltage), proceed as follows:
(A-B) Extract the fuse holder drawer from the power module.
(C) Insert the fuses making reference to the following chart:
Mains Voltage
Mains Voltage
(D) Extract and rotate the detachable part in way to read the correct voltage in the (E) window –
reinsert the fuse holder in the module.

6.6 CHECKING OF THE EQUIPMENT BEFORE EACH USE

Each time the use of the electrosurgical equipment is planned a check of the most important safety
aspects has to be implemented considering at least the following:
Check the integrity of cords, connections, wires breakage, etc.
Assure that all the electrical equipment is properly grounded.
Assure that all the accessories that should be used are available and sterilised.
Check, by disconnecting the reference electrode cable, the functioning of the OC light and
sound alarm warning.
§
Check, by activating the CUT and COAG power switch, the functioning of the emission
sounds warnings.

6.7 FUNCTION AND SAFETY CHECK AND TEST

At least once a year, the biomedical engineering department or other qualified personnel should do
the following check and test:
Check of the connectors and mains supply cord conditions.
Visual check of the mechanical protections.
Manuale d'Istruzioni / Instruction's Manual / Manual de Instrucciones
DIATERMO 108
Probable Cause
Error of the reference voltage value
Error in the output power activation
circuit
115 Vac Delayed fuse T4A - 5 x 20 mm
230 Vac Delayed fuse T2A - 5 x 20 mm
Solution
Verify the main voltage
Call for Service
Call for Service
MA254IGBEa
44