INSTALLATION & SETUP
P P R R E E H H E E A A T T S S T T A A N N D D B B Y Y S S E E T T U U P P
Preheat and Standby mode are two individual modes accessed
by a common button. When either the Preheat or Standby
Function is activated or active, the blue indicator light will
illuminate.
P P R R E E H H E E A A T T : : Upon initial Fluido DHT power up, the desired Preheat
mode can be activated to quickly bring the Cellex media to
clinical temperatures. Activation of preheat is accomplished by
pressing the Preheat Standby button, the Preheat mode operates
the Fluido DHT at 50% airspeed for 30 minutes or until the
Preheat Temperature is reached. Preheat can be automatically
timed to activate and is reviewed in detail in Preheat Standby
Start Time section.
S S T T A A N N D D B B Y Y : : The Standby mode is automatically entered at the
end of the Preheat cycle or when a timed therapy session is
completed. Standby operates the Fluido DHT for a period of two
hours to keep the Cellex media within ±5 °F (±3 °C) of setpoint.
To turn off the Standby mode, simply press the Preheat/Standby
button once and the blue indicator light will turn off. The unit
then reverts to a three minute cool down period before it
automatically shuts off.
Programming of the unit for the Preheat Standby function hours
per day and number of days per week is accomplished by
performing the steps detailed in Preheat Standby Start Time.
N N O O T T E E : : After using the unit for the first time every day, Standby
Mode does not need to be activated unless the time between
treatments is over 3 hours. The time to reach clinical
temperatures from Standby Mode is approximately 10 minutes.
1 1 2 2
Fluido DHT
™
Dry Heat Therapy Unit
P P R R E E H H E E A A T T S S T T A A N N D D B B Y Y S S T T A A R R T T T T I I M M E E
Press and hold the PREHEAT STANDBY
button for approximately five seconds.
This will cause the Pre-Heat Start Hours
to flash.
Use Up Arrow or Down Arrow button to
set the desired hour.
Press and release the PREHEAT
STANDBY button. This will cause the
minutes to flash.
Use the Up Arrow or Down Arrow
button to set the desired minutes.
Press and release the PREHEAT
STANDBY button.for AM/PM selection.
Use the Arrow Up or Arrow Down
button to set AM or PM.
P P R R E E H H E E A A T T S S T T A A N N D D B B Y Y T T I I M M E E
Press and release the PREHEAT
STANDBY button. This will cause the
Standby Hours to flash.
Use Up Arrow or Down Arrow button to
set the desired hours per day that the
unit is to maintain the Standby default
temperature.