Descargar Imprimir esta página

Ca-Mi New Askir 118 Basic Manual De Instrucciones página 19

Ocultar thumbs Ver también para New Askir 118 Basic:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

disinfectants by carefully following the instructions and dilution values supplied by the manufacturer.
After cleaning, leave the parts to dry in an open, clean environment.
Dispose of the aspiration catheter according to that provided by local laws and regulations.
The silicone aspiration tubes and the conical fitting may be carefully washed in hot water (temperature must not exceed
60°C). After cleaning, leave the parts to dry in an open, clean environment.
When cleaning is complete, reassemble the container for liquid aspirations according to the following procedure:
Place the overflow valve into its seat in the cover (under VACUUM connector)
Insert floating valve keeping the o-ring towards the opening of the cage
Place the o-ring into its seat around the cover
After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leackages or
liquid exit
After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly.
Then soak in warm water (temperature shall not exceed 60°C). Wash thouroughly and if necessary use a non-abrasive brush to
remove incrustations. Rinse in running warm water and dry all parts with a soft cloth (non-abrasive).
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C
(1 bar relative pressure – 15 min) making sure that the jar is positioned upsidedown. Mechanical resistance of the jar is
guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1). Beyond this limit the
physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended.
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged.
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min).
The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min).
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
Instruction for disposal Liner Flovac®:
If the device is equipped with disposable collection systems FLOVAC ® carry out the disposal of the bag as follows:
Turn off the Vacuum and remove all the tubes connected to the Liner, giving particular attention to avoid accidental contamination.
Fit the appropriate plugs to the "PATIENT" and "TANDEM" ports, pressing the home firmly, taking care to avoid accidental
contamination. Remove the liner bag from the rigid container and transfer it to the waste disposal area, ensuring that all the
openings are sealed, keeping in mind the product is potentially infectious. This product must be disposed of in accordance with the
current hospital regulations.
The NEW ASKIR 118 BASIC suction equipment does not need maintenance or lubrication.
It is, however to inspect the unit before each use. With regard to training, given the information contained in the user manual and
since it is easy to understand the said device, it doesn't appear to be necessary. Unpack the instrument and always check integrity
of plastic parts and AC/DC switching adapter , feeding cable, they might have been damaged during previous use. Connect cable to
electrical network and turn switch on. Close the aspirator outlet with your finger and with suction regulator in maximum vacuum
position check that the vacuum indicators reaches – 75kPa (-0.75 bar). Rotate the knob from right to left and check the aspiration
regulating control. Verify that loud noises are not present, these can indicate wrong functioning.
The device is protected by a safety fuse (F 10A L 250V) situated in the cigarette lighter cable. When replacing, always check the
type and value as indicated. Internally, the device is protected (see electrical specifications) by two fuses (T 15A L 125V) that
cannot be reached from the outside. Therefore, contact the manufacturer to request the assistance of an authorized and qualified
technician when they need to be replaced. If it's replaced make sure that its replacement is always the same type and value, as
indicated. The device is equipped with a Lithium-Ion Battery, which cannot be reached from the outside.
Only refer to CA-MI technical service personnel to replace it.
The device should be checked at least once every 12 months by technical service.
Every 24 months it is compulsory to have a safety inspection and technical maintenance performed on the device.
Fault type
1. Fixed Red Led
2. No light
3. No aspiration
4. No aspiration
5. The float doesn't close
6. The float doesn't close
7. Low suction
8. No aspiration due to flow
leakage of mucus
9. The Vacuum power on the
patient side is either very
PERIODICAL MAINTENANCE CHECKS
Cause
Battery run down
Defective AC/DC adapter or
technical internal problem
Jar Cap badly screwed down
Lid seal not in its seat
If the cap has been washed,
ensure that the float is not
partially detached
The float it's covered by dirty
material
Foam inside the jar
Filter blocked
 Protection filter blocked or
damaged
Hook up the power cord to the electricity mains,
positioning the equipment power switch on 0.
Contact the technical service
Unscrewed the cap, then rescrew it correctly
Unscrew the cap and insert the seal properly in its seat
Insert the float into it's place
Unscrewed the cap, leave the and put in on autoclave
Fill the jar to 1/3 full of ordinary water
 Replace the filter
 Replace or reconnect the tubes, check the jar
19
Solution
Replace filter
GB

Publicidad

loading