Descargar Imprimir esta página

Ca-Mi New Askir 118 Basic Manual De Instrucciones página 24

Ocultar thumbs Ver también para New Askir 118 Basic:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

GB
The surgical aspirator NEW ASKIR 118 BASIC is intended for use in the electromagnetic environment specified below.
The customers or the user of the surgical aspirator NEW ASKIR 118 BASIC should assure that it's used in such an
Immunity Test
Conducted Immunity
EN 61000-4-6
Radiated Immunity
EN 61000-4-3
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by reflection from buildings, objects and people.
a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and terrestrial
mobile radio, amateur radio devices, radio AM and FM transmitters and TV transmitters can not be theoretically and
accurately foreseen. To establish an electro-magnetic environment generated by fixed RF transmitters, an electro-magnetic
study of the site should be considered. If the field intensity measured in the place where the device will be used surpasses the
above mentioned applicable level of conformity, the normal functioning of the device should be monitored. If abnormal
performance arises, additional measures such as changing the device's direction or positioning may be necessary.
b) The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 10 V/m.
Recommended separation distance between portable and mobile radio-communication devices and the monitor
The NEW ASKIR 118 BASIC surgical aspirator is intended to operate in an electro-magnetic environment where RF irradiated
interferences are under control. The client or operator of the NEW ASKIR 118 BASIC device can help prevent electro-magnetic
interference by keeping a minimum distance between the portable and mobile RF communication devices (transmitters) and
the NEW ASKIR 118 BASIC device, as recommended below, in relation to the radio-communication maximum output power.
Maximum nominal
output power of the
Transmitter W
0,01
0,1
1
10
100
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres
(m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output
power of the transmitter in Watt (W) depending on the transmitter's manufacturer.
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by the reflection from buildings, objects and people.
Guidance and manufacturer's declaration – Immunity Emissions
Level indicated by
Compliance
the EN 60601-1-2
3Vrms 150kHz to
V
80Mhz (for non life-
supporting devices)
10V/m 80MHz to
E
2.7GHz (for non life-
supporting devices)
Separation distance from the frequency transmitter (m)
150 kHz to 80 MHz
d = [3.5 / V
]
P
1
0,12
0,38
1,2
3,8
12
environment.
Electromagnetic environment - guidance
Level
= 3 V rms
The portable and mobile RF communication devices,
1
including cables, must not be used closer to the NEW
ASKIR 118 BASIC device, than the separation distance
calculated by the equation applicable to the
= 10 V / m
1
transmitter frequency. Recommended separation
distance
d = [3.5 / V
d = [12 / E
d = [23 / E
Where P is the maximum nominal output voltage of
the transmitter in Watt (W) depending on the
manufacturer of the transmitter and the recommended
separation distance in metres (m). The intensity of the
field from the fixed RF transmitters, as determined by
an electro-magnetic study of the site
than the level of conformity of each frequency interval
. It is possible to check for interference in proximity
b)
to devices identified by the following symbol:
80 MHz to 800 MHz
d = [12/E
]
P
1
0,12
0,38
1,2
3,8
12
24
]
P
1
]
from 80 MHz to 800MHz
P
1
]
from 800 MHz to 2.7 GHz
P
1
a)
800 MHz to 2.7 GHz
d = [23/E
0,23
0,73
2,3
7,3
23
, could be lower
]
P
1

Publicidad

loading