THE CUFFS ARE INFLATING IN THE WRONG ORDER
Make sure that the multi-plug is correctly inserted into the device. There is a small additional nipple, the multi-plug, which must be inserted into the hole
provided for this purpose. lf the multi-plug is inserted into the device the wrong way, the multi-plug will be placed crooked.
MY VENEN ENGEL IS VERY LOUD
One should be able to talk or relax and chat with another person, without any issues, during the application. lf this is not the case, due to the Venen Engel
being too noisy, it is an indication that the transport screw on the bottom of the device has not yet been removed. Open the cover on the bottom and remove
the safety screw with the enclosed screwdriver. The screw is only used for securing the de vice during transport, and it can subsequently be disposed of.
YOU CAN HEAR AIR ESCAPING FROM THE DEVICE, HOSES ,OR CUFFS
(A)
Check the hoses and plugs for damage.
(B)
Check that the multi-plug is correctly plugged into the unit.
(C)
Check if the hose is bent or has been pulled off.
23