FIREBOY eco – Operating instructions V00
3
Start-up
Purpose This chapter defines the requirements for the gas connection and the
electricity supply of FIREBOY eco.
3.1
Connecting FIREBOY eco to the electricity supply
Insert the mains plug in the socket on the reverse side of the device (see
"Rear view" on page
3.2
Connecting FIREBOY eco to the gas supply
FIREBOY eco is appropriate for use with natural gas (gas mixture, >85%
methane, further alkanes) and commercial butane/propane gas (mixture
65%-10% butane, 20%-40% propane).
3.2.1 Fitting the nozzle for the gas in use
The P60 nozzle for butane/propane gas is fitted ex-factory for FIREBOY eco.
If necessary, the nozzle must be fitted for the gas in use.
Type of gas
Natural gas (methane)
Butane/propane gas
The nozzles are marked on the lower surface with the appropriate
designation.
C
AUTION
FIREBOY eco may only be operated with the nozzle that corresponds to the
gas type in use. Only use the nozzles supplied. Prior to every nozzle
installation, inspect the O-ring in the brass screw for damage and if
necessary replace it.
www.integra-biosciences.com
6) and connect to the power source.
Gas input pressure
20 mbar
50 mbar
Start-up
Inscription nozzle
N80
P60
7