FIREBOY eco – Operating instructions V00
1.1.3 Hazard warnings and caution notes of the red LED
R
LED
ED
Error.
The warnings are explained in detail in Chapter 4, see
page
13.
1.2
Appropriate use
The FIREBOY eco is an automatic safety Bunsen burner for heating or
flame-sterilizing suitable laboratory materials. It must not be used in
continuous operation for heating purposes.
FIREBOY eco may not be altered in design or in any of its safety aspects. If
alterations are made, INTEGRA Biosciences will be relieved from all liability
for any resulting damage.
1.3
Safety features of the FIREBOY eco
The FIREBOY eco is equipped with the following safety devices:
• If the flame has not ignited after about 3 seconds, the flame monitor
closes off the gas valve.
• If the flame is inadvertently extinguished, an attempt will be made
to reignite it. If this is not successful, a message is displayed.
• The burner chamber and the valve housing are monitored by a
temperature sensor. If the temperature exceeds the safety level,
the gas valve is automatically closed off.
2
IS FLASHING OR ILLUMINATED
Introduction
"Troubleshooting" on