6. Close the window once all upgrades have been completed.
Manually Initiating Calibration/Reconditioning
ATTENTION: Manual calibration or reconditioning does not
apply to IMPRES or IMPRES 2 batteries that has this icon.
Though Calibration/Reconditioning is automatic, there may be situations
in which manual initiation is desired. For example, the IMPRES 2 or
IMPRES charger may be indicating Alternating Amber/ Green for the
IMPRES 2 or IMPRES battery, because Calibration is disabled in the
charger. If within 30 minutes that battery is moved to an IMPRES 2 or
IMPRES charger that has Calibration enabled, the last charging state
will resume, preventing automatic start of Calibration/ Reconditioning.
To manually initiate Calibration/Reconditioning, ensure that the
IMPRES 2 or IMPRES battery is removed from the charger for more
than 2-1/2 minutes. Then, perform the following steps:
1. Insert the battery into the charger pocket
2. Within 2-1/2 minutes, remove the battery from the charger pocket.
3. Within 5 seconds, reinsert the battery into the charger pocket.
Calibration/Reconditioning starts immediately, typically beginning with
Battery Discharge (Steady Amber). Calibration/Reconditioning is
complete only after full charge (Steady Green).
11