Error Messages - Xylem SI Analytics HandyLab 700 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Error Codes and Device Messages

Error Messages

The following error messages can be shown in the display.
Message Cause
Battery empty
blinks
ERROR 1
pH value out of range
ERROR 2
ORP value out of range
Temperature value out of
ERROR 3
range
Sensor zero point
ERROR 4
too high/low
ERROR 5
Sensor slope too high/low
Calibration error:
ERROR 8
Identical buffers
Calibration error:
ERROR 9
Buffer unknown
ERROR
Cal media interchanged
10
ERROR
Measured value unstable
11
Drift too high
ERROR
Time and date invalid
14
ERROR
Confi guration invalid
18
ERROR
Factory settings error
19
ERROR
Sensor error (Memosens)
21
Remedy
Replace batteries
Check whether the measurement
conditions correspond to the ad-
justed measuring range.
Thoroughly rinse the sensor and
recalibrate. If this does not help,
replace the sensor.
Use a buffer solution with a diffe-
rent nominal value before starting
the next calibration step.
Make sure that you use the same
buffer set as confi gured.
Repeat calibration.
Leave the sensor in the liquid until
the temperature is stable. If this
does not help, replace the sensor.
Set time and date
Restart,reset to factory settings
(Setup: DEFAULT YES), confi gure
and calibrate.
If this does not help, send in the
device for repair.
Device defective, send it in.
Connect operational Memosens
sensor.
35

Publicidad

Tabla de contenido
loading

Tabla de contenido