Assembly
4
Assembly
4.1
Dimension drawing of the mounting plate
Fig. 4
The dimension drawing is to be used to mark the 4 drilling holes for the mounting plate.
NOTICE
Damage to the device
If the device holder is not fixed tightly enough, it may loosen and fall.
► Always mount the ceiling holder and the mounting plate with the 4 screws provided.
► The assembly must be performed by specialized personnel only.
verify sufficient sturdiness. For further information, see the "Technical data" chapter.
Distance between the holes on the mounting plate: 95 x 95 mm
4.2
Mounting the wall holder
1. Make sure that the surface is sturdy enough.
2. Drill the 4 fixation holes according to the dimension drawing.
3. Screw the mounting plate to the wall with 4 M6 screws.
4. Screw the wall holder to the mounting plate with the 4 screws.
16
Dimension drawing of the mounting plate (measurements in mm)
Before mounting, check the surface, onto which the device holder is to be mounted, to
95
|
Instructions for use
Device holder for Oxylog VE300
95