6. Commissioning, operation
Function locking
The calibration system model BCS10 is delivered from the factory with unlocked access
to all settings and they can be changed.
The access to the adjustable parameters can be limited or completely prohibited to
GB
prevent unauthorized changes of the configuration.
This is achieved by the function locking.
In addition, user defined units can be scaled here.
For configuration of the function locking see chapter 6.7 "Configuration of the function
locking".
6.7 Configuration of the function locking
Scroll to the menu item "FUnC
LOCK
" and press the ENTER key. On the digital indicator
appears the message "0PWRD". Enter the password "101". The password is factory set
and cannot be changed.
The functions that can be locked or unlocked:
- zero point adjustment (ZERO button)
- setting the units
- auto Power-Off Settings
- setting of the damping
- setting of the sampling rate
- tare setting
The command LOC locks the respective function and UnLOC unlocks it again.
Locked functions are not displayed in the configuration menu any more and therefore
cannot be changed.
Setting of a user-defined unit or a user-defined measuring range:
In the menu item SET FACTOR, a multiplier between 0.001 and 100 can be set for
a user-defined measuring range. The defined factor is multiplied by the measured
pressure value in psi. The result is displayed.
Example: 40 psi correspond to 1.000 g of a product in a tank. The weight of the product
should appear on the display.
If 25 is defined as the multiplicator, the pressure of 40 psi corresponds to the value
1.000 (40 x 25). „CUST" is displayed on the instrument as a technical unit.
20
WIKA operating instructions calibration system, model BCS10