Descargar Imprimir esta página

WIKA BCS10 Manual De Instrucciones página 24

Ocultar thumbs Ver también para BCS10:

Publicidad

Idiomas disponibles

Idiomas disponibles

6. Commissioning, operation
GB
24
Compare and record the displayed value of the test
item with the displayed value of the digital pressure
gauge.
To improve the readability of the display, the backlighting can be activated
by means of the ENTER key.
Open the pressure relief valve after calibration to
depressurize the system.
The test item can be separated from the calibration
system.
The calibration system can be dismount now in the
reverse order.
WARNING!
Only disconnect once the system has been depressurised!
WIKA operating instructions calibration system, model BCS10

Publicidad

loading