Disinfect the Scanner
After each patient, the scanner must be thoroughly disinfected.
To adequately disinfect the scanner, follow the disinfectant manufacturer's instructions for the
appropriate contact time.
CAUTION: If the scanner is visibly soiled, it must be thoroughly cleaned prior to disinfecting. See "Clean
the Scanner"
To disinfect the scanner, follow these steps:
• Remove the Scanner Tip.
• Remove all visible soil (see "Clean the Scanner").
• Use a commercially prepared intermediate level disinfectant wipe. Follow the manufacturer's
instructions for contact time.
Recommended disinfectant wipes: CaviWipes™*
WARNING: Using a disinfectant that has not been approved may cause damage to the scanner.
• Thoroughly wipe all surfaces of the scanner, DO NOT allow liquid to enter through the gap, air outlet,
or pin holes.
WARNING: Do not rinse.
• Allow to air dry.
• After the scanner has dried, use a clean, lint-free cloth dampened with water to remove residual
disinfectant from the surface of the scanner.
Clean and sterilize the scanner tip
WARNING
• Wear gloves when handling a contaminated Scanner Tip.
• Read and follow the warnings and personal protection instructions provided in the manufacturer's
SDS for the detergent used to clean the Scanner Tip.
• Do not soak the Scanner Tip in disinfectant for a long period.
• Dry the Scanner Tip thoroughly before mounting onto the scanner.
• Do not use an ultrasonic cleaning machine to clean the Scanner Tip.
• Do not soak the Scanner Tip with alcohol-based disinfectants.
Clean the Scanner Tip
To manually cleaning the Scanner Tip, follow these steps:
• Rinse excess soil from the Scanner Tip (2 minutes).
• Using a soft brush, apply enzymatic detergent solution (e.g., Metrex EmPower) to all surfaces.
• Rinse under clean, running water (2 minutes).
• Inspect Scanner Tip. If the Scanner Tip is not clean, repeat the last three steps.
• Use a lens tissue or lint-free cloth to remove any dust from the mirror in the Scanner Tip.
* CaviWipes are a trademark of Metrex Research, LLC.
NeoScan™ 2000 Installation and operation manual // Maintenance