Operation
Auxiliary display for check and reference weighing
Net indicator
[ON/OFF]
Measured value
Sign
Measuring unit
Unreliability
indicator
Info line
Function in an
Application (Soft key)
Function in a Menü
(Navigation)
[Left]
[Right]
[Up]
[Down]
[PRINT]
[MENU]/[ESC]
[0/T]
[Enter]
[SWITCH]
The measured value is stable as soon as the unit, e.g. "g", appears. The unreliability indicator lights
up while the scale is recalculating the reference weight, for example, during counting, or when it is
verified and in the warm-up phase.
Application example: Adding up components (application net total)
If the scale is not set to English, enter the configuration menu: Switch the scale off and on with
[ON/OFF] and, as soon as all segments of the display light up, press and hold [MENU] until
"UNIT-1 g" appears. Press [Down] repeatedly until "SPRACHE DEUTSCH/LANGUE FRANÇAISE"
appears. Press [Enter]; "SPRACHE DEUTSCH/LANGUE FRANÇAISE" starts flashing. Press [Down]
repeatedly until "LANGUAGE ENGLISH" appears and confirm with [Enter]; "LANGUAGE ENGLISH"
stops flashing. Press [ESC]; the scale finishes starting up.
To enter the application menu, press [MENU]; "SET APP. OFF" appears. Press [Enter]; "OFF" starts
flashing. Press [Down] repeatedly until "NET-TOTAL" appears and confirm with [Enter];
"NET-TOTAL" stops flashing. Press [ESC]; the scale exits the application menu, starts the
application Net total and displays 0g as the measured value and "STO 0
RES INF" in the info
line.
Place the first component, e.g. 1000g, and press {STO}; the scale records the value, tares itself
and displays "STO 1
RES INF" in the info line. Place the second component, e.g. 2000g, and
press {STO}, and finally place the third component, e.g. 200g, and press {STO}.
While adding up, the current total weight can be viewed at any time by pressing [SWITCH]. The
scale then switches to weighing mode. To return to the application, press [SWITCH] again. To
switch from the analogue display to the application commands in the info line, press [SWITCH]
twice; the scale switches to the weighing mode and back to the application. It then displays e.g.
"STO 3
RES INF".
Press {INF}; the scale displays in the info line all information on the weights just added up. Use
[Down] and [Up] to scroll through the entire stack of info lines: Total, remaining capacity and all
individual weights, e.g. 1000g, 2000g and 200g. Press [ESC] to exit the information query; the
scale switches back to the running application and is ready for another weighing-in.
To restart, press {RES}; the scale resets the counter to 0.
To quit the application, press [SWITCH]; the scale switches back to weighing mode.