3. Loading (solid instruments).
Place UNWRAPPED and WASHED instruments
ONLY plus a TST Strip into the instrument basket
or cassettes, and place in the unit.
If using a basket, place the metal 'V' support in
the bottom of the unit so ensuring instruments
and TST Strips are not in the water .
TST Strips must be placed as near to the centre of
the load as possible.
4. Closing.
Always place the lid on the autoclave with the
Depressurization Valve (A) in the "open"
position. Align the arrows on the lid and body
(S2) and turn in a clockwise direction ensuring it
is completely closed (S3).
Close the Depressurization Valve (A) so its arrow
is in line with the "O" on the lid decal.
5. Power on.
Switch the unit's power on by pressing Rocker
Switch (B). (S4)
Lights: L2 & L7 illuminate GREEN
6. Temperature.
The 134˚C temperature is automatically selected
when the power is turned on.
Select 121˚C temperature by pressing button (C).
Either the 134˚C or the 121˚C indicators will be
illuminated (S5).
7. Starting.
Depress the lid interlock button (D) to start the
sterilizing cycle. (S6)
Lights: L6 & L7 illuminate GREEN
L2 illuminates ORANGE
L5 MAY flash ORANGE
- as the temperature rises, air will be displaced by
steam through the Air Bleed Device located in
the lid, until it closes with an audible
"click", sealing the unit.
-Sterilizing Temperature is reached when:
Lights:L6 & L7 illuminate GREEN
L2 & L4 illuminate ORANGE
L5 flashes ORANGE
-Sterilizing Cycle is completed when:
The internal Buzzer sounds.
Lights: L3 & L7 illuminate GREEN
L2 illuminates ORANGE
8. Depressurizing.
Once the Sterilizing Cycle has been completed, the
unit needs to be depressurized and allowed to cool
down before the lid and sterilized instruments can
be removed.
The time taken to reach the point at which this is
safe to do so can be shortened by Manually
Depressurizing the unit.
Open the depressurization Valve (A) by turning
slowly in an anti-clockwise direction.
Lights:L2 illuminates ORANGE
L3 & L7 illuminates GREEN
Warning: There will be a visual and audible
release of steam from the rear of the lid.
9. Unlocking.
Once the temperature has reached a safe level an
internal buzzer sounds, indicating the lid can be
unlocked.
Lights:L7 illuminates GREEN
L2 flashes GREEN
Unlock lid by pressing and releasing button (D).
Lights: L7 & L2 illuminate GREEN
Remove the lid by turning in an anti-clockwise
direction, aligning the arrows (S2 & S3).
10. Unloading.
Lift off the lid, gently place upside down on a
solid work surface and leave to cool. Ensure the
Depressurization Valve (A) is in the closed
position to avoid damaging it.
The unit has completed a successful cycle if the
"spot" on the TST Indicator Strip has completely
changed colour from yellow to purple.* The
basket or cassettes containing the sterilized
instruments can now be lifted out of the unit.
Allow 5 minutes for the lid to cool down before
replacing as per Step "4".
*Please Note. If the "spot" has not completely
changed colour, replace with a new TST strip and
start a new cycle. If the "spot" fails to change
colour for a second time, do not use the unit until
checked by a qualified engineer.
DO NOT USE THE INSTRUMENTS IF A
COMPLETE STERILIZIATION CYCLE HAS
NOT BEEN ACHIEVED.