In the event of a fault occurring during any stage of the unit's operation, identify the fault by referring
to their descriptions below. The fault can be corrected by following a few easy steps.
(Note: light L7 will be constantly illuminated GREEN whilst the unit is powered up).
FAULT MODE DESCRIPTION
Fault 1: Failure to complete a satisfactory
Cycle
Lights: L1 flashes RED
L2 illuminate ORANGE
The buzzer sounds continuously after 5
minutes
Fault 2: Low Water or Boil Dry
Lights: L1 flashes RED
L2 illuminates ORANGE
L3 flashes RED
Buzzer sounds continuously
Fault 3: Temperature Sensor Warning
Lights: L1 flashes RED
L2 illuminates ORANGE
L4 flashes RED
Buzzer sounds continuously
Fault 4: Lid Interlock Shaft Over-travelled
Lights: L1 flashes RED
L2 illuminates GREEN
Fault 5: Steam or water leaks from under the
lid
Fault 6: No power to unit
Lights: All lights fail to illuminate
FAULT MODE REMEDY
Important - Before restarting cycle check the
mains lead is firmly connected to the unit and an
EARTHED/GROUNDED power supply.
Steps 1-4 should be followed for ALL Fault
modes experienced.
1. If the buzzer is sounding constantly, silence
by pressing Reset Button (E).
All lights flashing RED will change to RED
2. Switch the unit's power "OFF" then "ON" -
Rocker Switch (B).
Lights: L1 flashes RED
L2 illuminates ORANGE
Troubleshooting
3. Press Reset Button (E)
- if the Temperature is above boiling point
L2 illuminates ORANGE
- if the temperature is below boiling point
L2 flashes GREEN
IF ABOVE BOILING POINT WAIT FOR
UNIT TO COOL
4. When the temperature is below boiling point:-
Press Cycle Start button (D) to return unit
to ready state. The lid may now be
removed. If the lid is not on, manually
push the Lid Interlock Shaft down to
revert to ready state.
5. Further action can now be taken
depending on the fault mode experienced
Faults 1 and 4
No other action necessary, restart cycle. If
fault repeats arrange for a service engineer to
visit.
Fault 2
Fill to the waterline with distilled or
de-ionised water - restart cycle.
Fault 3
Call Service Engineer. DO NOT USE UNIT.
Fault 5
Wash gasket as described under "Care &
Maintenance", restart cycle. If fault persists,
replace with a new gasket.
Fault 6
Check for blown fuse, defective socket, mains
not connected. If fault discovered and can be
remedied do so, then restart cycle. Otherwise
arrange for a service engineer to visit.
Should all power be lost during a cycle, the lid
cannot be removed until power has been restored.
Should there be an internal power failure during a
cycle, the lid cannot be removed until the fault
has been rectified by a service engineer.
7
Lights: L1 illuminates RED
Lights: L1 illuminates RED
L2 flashes GREEN
GB