Care and cleaning
Braun ThermoScan probe covers are intended for single use only!
To ensure accurate readings, always use Braun ThermoScan
disposable probe covers with the Pro 3000 thermometer.
Damaged, perforated, or soiled probe covers can lead to
inaccurate temperature readings. Additionally, the accuracy of
your temperature readings can be affected by damage to the
probe window, or the presence of dirt and earwax on the probe
window.
Probe Window
To assure a high level of accuracy, it is very important to keep this
window clean, dry and undamaged. Fingerprints, earwax, dust
and other soiling compounds reduce transparency of the window
and result in lower temperature readings. If the thermometer is
accidentaily placed in the ear without a probe cover installed,
the window must be cleaned immediately, Additionally, the window
and/or probe tip area should never be modified, changed, or
adjusted. These changes will affect calibration and accuracy of
the thermometer.
To clean the window, gently wipe its surface with a cotton swab
slightly moistened with alcohol and immediately wipe dry with a
clean catton swab. After cleaning, allow at least 5 minutes drying
time before taking temperatures. Make sure window is clean and
dry. Avoid touching the window except when cleaning is required,
If the window is damaged, the thermometer must be returned to
Braun ThermoScan for service.
Note:
Do not use any chemical other than alcohol to clean
the probe window.
Thermometer
Use a soft cloth slightly moistened with alcohol to clean the
thermometer display and exterior. Do not use abrasive cleaners.
Never submerge the thermometer in water or any other liquid.
Do not autoclave or immerse. Do nat use ammonia or ammonium
chloride based agents as they could damage the exterior of
the instrument. If you currently use a specific cleaning agent or
disinfectant, we recommend that you examine its chemical
ingredients prior to use on the thermometer.
Store thermometer and probe covers in a dry location free from dust
and contamination and away from direct sunlight. The ambient
temperature at the storage location should remain fairly constant
and within the range of 50°F to 104°F (10°C - 40°C).