After the measurement is completed, the temperature value measured
remains in the display until the SC 1091 is switched off. To turn the
device off after use, press the 0/II button briefly. The signal tone will again
be sounded. If the device is not switched off, a signal tone is sounded
after around 3 minutes and the device then switches off automatically.
Function test
The thermometer is checked for correct functioning automatically every
time it is switched on. All segments are displayed, and then L°C. If a
failure (measurement inaccuracy) is recognised, "ERR" (Error) is shown
in the display.
Display of last measured value
If you keep the 0/I button pressed for longer than three seconds, the last
measured value will be shown in the display. As soon as you release the
0/I button, the device is ready to take measurements.
Cleaning and disinfection
The veterinary thermometer is watertight.
Please note that watertight does not mean that the thermometer
is also tight in aggressive disinfection or solvent solutions!
The thermometer can be cleaned by wiping with any of the disinfection
tissues listed below or with a moist cloth containing any of the following
disinfection agents.
Disinfectant
Supplier
®
Kodan
tissues
Schülke & Mayr
®
EuroSept
Max
Henry Schein Inc.
Wipes
Medizid Rapid
Rösner-Maut by Meditrade
GmbH
®
Helipur
H plus N
B. Braun Melsungen
Changing the battery
As soon as the
symbol appears in the display, the battery has run out
and requires replacement. Always replace the discharged battery with
one of the same type. To do this, proceed as follows:
Pull the cover off the battery compartment. The battery can now be
accessed. Remove the used battery with the help of a pointed object
(pincers, small screwdriver etc.) and put the new battery (LR 41 or SR41)
in. Please make sure it is the right way round.
Guarantee
We give a guarantee period of 2 years from the date of purchase. We
Use
According to supplier
instructions for plastic
surfaces
- 3 -
Special features
The SC 1091 is a maximum thermometer, i.e. the highest temperature is
displayed latest after 25 seconds. The most recent measured value is
always saved.
Safety information
♦ The SC 1091 is suitable only for measuring animal body temperatures.
♦ The thermometer contains small parts (batteries, etc) which could be
swallowed by small children. Please keep the thermometer and its
batteries out of the reach of small children.
♦ Protect the SC 1091 from direct sunshine and temperatures above
50°C.
♦ Never drop the SC 1091 or open it (with the exception of the battery
compartment for battery replacement).
Operating manual
Press the 0/I button to switch on. A short signal tone shows that the
device is in the "switched on" condition. A segment test takes place
simultaneously (checks all display segments and carries out an internal
function test (see under "Function test")).
check all display
segments
After this, if the temperature measured by the temperature sensor is less
than 32°C, the display shows an "L" and "°C". After around one second,
the "°C" starts to blink. The „L" stands for low (less than 32°C).
The current temperature is continuously displayed during measurement,
and the "°C" blinks during the entire measurement time. The
measurement is completed after around 25 seconds.
guarantee perfect design of this article. We will rectify all material or
manufacturing errors during the guarantee period. The guarantee
excludes damages caused by improper handling, related to the battery or
to the packaging.
Specifications
Type:
maximum thermometer
Measuring range:
32.0°C to 43.9°C
Temperature below 32.0°C: display "L" for low.
Temperature above 43.9°C: display "H" for high.
Measurement accuracy:
+/- 0.1°C between 35.5°C and 42.0°C at a
ambient temperature of 18°C to 28°C.
Self-test:
Automatic internal check of test value 37.0°C. In
case of deviation > 0.1°C, "Err" will be displayed.
Display:
3-digit liquid-crystal (LCD display).
Battery type:
1 no.: LR 41 or SR 41.
Battery display:
The
Storage temperature:
0°C to 50°C
Disposal note
Defect devices or devices which are taken out of operation at the
end of their service life may still contain residues which could
present a risk to the environment. Please dispose of the device in
a suitable and correct way at a collection point for old electrical
items.
Batteries must be disposed of correctly. There are containers for
battery disposal at battery dealers and in communal collection
points.
Explanation of symbols
ready for the
measurement
- 2 -
shows when the battery is discharged.
Scala Electronic GmbH
Ruhlsdorfer Straße 95
14532 Stahnsdorf
www.scala-electronic.com
Rev. 31/05/2022/GB