5. Cleaning
It is important to follow the cleaning procedures to avoid cross contamination. Be sure to clean
the surface in a dry and dust free environment. Wipe down the pump unit with a damp cloth
pre-soaked with a mild detergent. Avoid contact with dust and proximity to dusty areas. Make
sure that any cleaning agents you use will not harm or corrode the plastic casing on the pump
unit. If your doctor or medical facilities have other special cleaning instruction, please follow the
professional instruction.
CAUTION: Do not immerse or soak pump unit.
WARNING: Do not remove the casing of the pump to avoid electrical
shock. All disassembly or repair should be done by authorized
technicians.
CAUTION: The pump does not need oil lubrication; please do not
dissemble the system.
Cover Material: Stretch
Silver
Wipe down the mattress with 70 % alcohol solution or chlorine solution concentration must be
less than or equal to 1,000 ppm. After disinfecting the mattress, use water and a soft cloth to
wipe off all traces of disinfectant and make sure the mattress is dry before storage. Also the
mattress top cover can be completely removed for laundry with water temperature indicated on
the laundry label; however, it is recommended that user should check with local policy or
regulation to determine the time/temperature ratio required achieving thermal disinfection. The
cover may also be cleaned using sodium hypochlorite diluted in water. After cleaning, please
avoid dust and proximity to dusty areas and all parts should be air dried thoroughly before use.
CAUTION: Do not use phenolic based products for cleaning.
CAUTION: After cleaning, dry the mattress without direct exposure of
sunlight.
CAUTION: Do not use ozone or UV disinfection as these methods can
damage the surface coating.
6. Storage
(1).
To store the mattress, lay the mattress out flat and upsides down.
(2).
Roll from the head end towards the foot end with CPR valve open.
(3).
Foot-end strap can then be stretched around the rolled mattress to prevent unrolling.
NOTE: Do not kink, crease or stack the mattresses and do not store the
system in direct sunlight, high temperature or moisture area.
+
Stretch
11