Optima Turn/English
(9). Cycle Time
 By pressing this button after changing the operation mode, Alternate Mode Cycle
Time can be selected between 10, 15, 20 and 25 minutes.
 By pressing this button after changing to the turning mode, Turning Dwell Time can be
selected between 10, 20, 30 and 60 minutes.
NOTE: Cycle time and dwell time duration can be selected based upon
patient comfort and desired outcome.
(10). Alarm Mute Key
When there is an alarmed situation happened, the number display will show corresponding
error code with buzzer sounded every few seconds. The buzzer can be temporarily
suspended by pressing the Alarm Mute Key. Alarm buzzer will signal repeatedly every 3
minutes until the alarmed issue is solved.
A. PFA (Power Failure Alarm)
The LED light and buzzer will activate when the power failure happens. To
disable both buzzer and LED, press the mute button.
NOTE: The pump can seal the air inside the mattress, which temporaily
keeps the patients from bottoming out during power failure. Patients
should be moved if the power failure issue cannot be solved soon.
B. Low Pressure Indicator
When low pressure LED lights up, the pressure inside the air mattress is below
normal range. Please refer to troubleshooting guide.
NOTE: The pump can seal the air inside the mattress, which temporaily
keeps the patients from bottoming out during low pressure condition.
Patient should be moved if the power failure issue cannot be solved
soon.
C. Tech. Support
The Service indicator light will activate when an issue occurs related to the
system. Please contact the local representative or place of purchase for
technical support.