sending the product to the manufacturer for service. A known calibration weight (in
kilograms) is required to recalibrate the scale, and can be any value between 5 kg and 100
kg. 25 kg is recommend. Best accuracy is obtained by using the largest weight available (i.e.
a 50 kg calibration weight will provide better calibration accuracy than a 5 kg calibration
weight). In addition, it is necessary to know the weight to within 2 grams.
To recalibrate the OCTA-WIRELESS, perform the following steps:
1. Place the OCTA-WIRELESS weighing platform on a firm, level surface.
2. Turn the power to the OCTA-WIRELESS off.
3. Remove all objects from the platform.
4. While holding
a few seconds the display will show SET CAL and 25.000 kg.
5. Using
or
weight of the reference calibration weight. Use the
6. Press
. A number representing the weight of the unloaded platform will appear on
the display. Press
7. Apply the calibration weight to the platform. A number representing the applied
reference weight will appear on the display. Allow at least 10 seconds for this number to
stabilize.
8. Press
. The OCTA-WIRELESS is now recalibrated to the reference weight.
9. Refer to the Quick Start section above for setting the desired Units and Resolution.
, press and release
adjust the displayed value (in 1g increments) so that it is equal to the
to lock in this value. The display will read zero.
OCTA-WIRELESS User's Guide
. The OCTA-WIRELESS will turn on, and after
key to adjust in 1 kg increments.