Descargar Imprimir esta página

Ki Mobility FOCUS PT Manual De Instrucciones Y Garantía página 13

Publicidad

Idiomas disponibles

Idiomas disponibles

IV. WHEELCHAIR USE & CONFIGURATION
D. Power Tilt (Continued)
Initial Charge
1. Identify the box with the charger.
2. Connect the charger cable to the extension cable (Fig. 2:A).
3. Plug the other end of the charger into a wall outlet for an initial charge.
4. Fully charge battery before first use.
Fig. 2
Charging
For more information on charging and the battery charger, refer to the battery charger owner's
manual that is included in the battery charger box.
The battery is charged through the external charger.
Charging is only allowed with charger provided.
Charging time for a completely discharged system is approximately 6 hours.
Charging must take place in a room with good air ventilation and a temperature of 0-40°C
(32-104°F).
Operation is prevented by the system during charging.
If stored, the battery must be charged before use.
Manual Tilt Change or Center of Gravity Adjustment
NOTE: A Center of Gravity adjustment is required for initial use and after any adjustment is
made to the chair.
1. With the occupant in the chair, ensure two people are present to assist in the Center of
Gravity adjustment.
2. One person must support the back of the chair to prevent a sudden drop while the other
person assisting pulls the motor disconnect pin (Fig. 3:A) on the power tilt assembly.
Fig. 3
A
A
12

Publicidad

loading