In addition, the TR8703 allows the user to store the data in automatic or
manual mode after each test eliminating the need to write down the test
results. Each measurement record is date and time stamped providing
complete test information.
The user can also save Nameplate voltages and compare the results as
the data is being gathered. The data can later be downloaded to a PC
and analyzed using the software package which is included with the
product.
The software allows full control of the instrument.
The TR8703 utilizes an advanced, low-voltage, step-down
measurement technique in which the high voltage "H" windings are
subjected to test excitation. This results in greater operator safety
and the ability to test a much wider array of transformer types and
sizes.