Hotplates & Stirrers
HP-200-C, ST-200-C, SHP-200-C,
SHP-200D-C, HP-200-S, ST-200-S,
SHP-200-S, SHP-200D-S
Introduction
Thank you for purchasing this Cole-Parmer
product. To get the best performance from the
equipment, and for your own safety, please read
these instructions carefully before use. Before
discarding the packaging check that all parts are
present and correct.
This equipment is designed to operate under the
following conditions:
- For indoor use only
- Use in a well ventilated area
- Ambient temperature range +5°C to +40°C
- Altitude to 2000m
- Relative humidity not exceeding 80%
- Mains supply fluctuations not exceeding 10% of
nominal
- Overvoltage category II IEC60364-4-443
- Pollution degree 2 IEC664
- Use with a minimum distance all round of
200mm from walls or other items
If the equipment is not used in the manner
described in this manual and with accessories other
than those recommended by Antylia Scientific, the
protection provided may be impaired.
Electrical Installation
THIS EQUIPMENT MUST BE EARTHED
Before connection please ensure that the line
supply corresponds to that shown on the
rating plate located on the base of the unit.
Power requirements:
Model
Wattage
HP-200-C
530W
ST-200-C
60W
SHP-200-C 600W
SHP-200D-C 600W
3
Model
Wattage
HP-200-S
750W
ST-200-S
60W
SHP-200-S
800W
SHP-200D-S 800W
There is a hot condition IEC socket at the rear of
the instrument for connection to the mains supply
(see figure 2). The unit is supplied with a mains
lead fitted with either a UK 3-pin or "Schuko"
2-pin hot condition IEC plug for connection to the
instrument.
Should the lead not be suitable for connecting to
the mains power supply, replace the plug with a
suitable alternative.
THIS OPERATION SHOULD ONLY BE
UNDERTAKEN BY A QUALIFIED ELECTRICIAN.
NOTE: Refer to the equipment rating plate to
ensure that the plug and fusing are suitable for
the voltage and wattage stated.
UK mains cable wiring is coloured as follows:
BROWN - LIVE
BLUE - NEUTRAL
GREEN/YELLOW - EARTH
US mains cable wiring is coloured as follows:
- BLACK - LIVE
- WHITE - NEUTRAL
- GREEN/YELLOW - EARTH
Should the mains lead need replacement, a cable
of 1mm
2
of harmonised code H05RR-F or
H05RN-F connected to an IEC hot condition plug
should be used.
IF IN DOUBT CONSULT A QUALIFIED
ELECTRICIAN