10.6
Restarting the device
Using the "Restart" menu function
196 • seca 655/654; seca 455; seca 257/256/254; seca 453
In individual cases (e.g. following implausible input) it may be necessary to
restart the device. To do so, use the
the power supply and then restore it.
NOTE
All the individual settings in the device are retained in the event of a
restart. If you want to reset the device to factory settings, proceed as
described in the relevant section:
1. Make sure that there is no load on the weighing platform.
2. Press the
key.
ð The
Settings
menu is displayed.
3. Press the
or
key until the
4. Press the
Restart
item.
5. Press the
symbol.
ð The device restarts.
6. Wait until the main screen is displayed again.
ð The device is ready for operation.
Restart
function in the menu or interrupt
è Factory settings, page 185
Restart
menu item is displayed.