9. Error Messages
Display
24
Display
Meaning
Measured
temperature
too high
Measured
temperature
too low
Ambient
temperature
too high
Ambient
temperature
too low
Malfunction
display (error)
Blank display
Dead battery
indication
Possible cause
and fault
correction
Thermometer displays
«H» when the measured
temperature is higher
than 42.2°C or 108.0°F.
Thermometer displays
«L» when the measured
temperature is lower
than 34.0°C (93.2°F).
Thermometer displays
«H» together with the
« » symbol when the
ambient temperature is
higher than 40.0 °C or
104.0 °F.
Thermometer displays
«L» together with the
« » symbol when the
ambient temperature is
lower than 16.0 °C or
60.8 °F.
Please contact your local
service centre or boso if
the display shows the
error sign consecutively.
Please check if the bat-
tery is in a good condi-
tion. Also check the po-
larity (<+> and <->) of the
battery.
The battery should be re-
placed immediately if the
battery icon is the only
symbol shown on
the display.